Personal Dosimeter Ra-D-60

Technical Specifications: Ra-D-60

Detector:Energy compensated Si-diode
Radiation:Gamma- and X-rays
PTB type approval:23.52 / 00.02
Measuring value:Hp(10)
Nominal working range:10 µSv ... 10 Sv
Measurement range:
Dose:1 µSv ... 9,999 Sv
Dose rate:*5 µSv/h ... 3 Sv/h
Measuring uncertainity:< ± 5 % (Cs-137, 662 keV, @ 2 mSv/h), Hp(10)
Energy response:60 keV ... 3 MeV, better than ± 25 %
Dose rate linearity:< ± 15 % (up to 3 Sv/h)

Acoustic alarms:

(at embedded speaker)

  • Dose, dose rate, dose overflow,
  • Dose rate overflow at 3 Sv/h,
  • Low battery,
  • Defect dosimeter
Sound level:> 85 dBA at 30 cm distance
Alarm thresholds:*6 presets, each for integrated dose and dose rate

Manually selectable by push button
Power Supply:1 triple A alkaline cell, (IEC LR03)

Life: 1800 h in background field (dose mode)
Push Button Functions:
  • Switch ON/OFF
  • Chirp ON/OFF
  • Change Alarm Thresholds
  • Activate Battery Test
  • Reset Integrated Dose
  • Switch: Dose/Dose Rate
Temperature range:Operating: - 20 °C up to + 50 °C

Storing: - 20 °C up to + 70 °C
Dimensions (H x L x W):78 mm x 67 mm x 22 mm
Weight:80 g incl. battery

* The dose rate display and the dose rate thresholds are not approved for calibration