Technichal Specifications OD-01 Hx

Survey Meter OD-01 Hx


Measured values:Photon equivalent dose Hx

Photon equivalent dose rate dHx / dt
Display ranges:
Dose rate:0 .. 2000 µSv/h
0 .. 2000 mSv/h
Dose:
0 .. 2000 µSv
Energy range:6 keV to 15 MeV (> 15 MeV with additional acrylic plastic shielding)
Radiation detector:Air-equivalent ionisation chamber
Volume:
600 cm3
Build-up cap:removeable, 550 mg/cm²
Intrinsic error:< 15 % (fine measurement range 20)
< 5 % (fine measurement ranges 200 and 2000)
Linearity:
5%
Saturation deficite:
5% @ 2000 mSv/h
Operating voltage control:Battery symbol in case of low battery voltage
Interface:USB (optionally with software)
Power supply:4 round batteries LR 06 type AA max. 6,2 V
6V DC external power supply optionally available
 Power consumption:
approx. 30 mA @ 6V
Battery operating life:approx. 100 hrs
Working temperature range:-10 °C to + 45 °C
Storage temperature range:-25 °C to + 55 °C
Ambient pressure:800 hPa to 1100 hPa
Relative humidity:< 80 %
Dimensions:
Display unit (L x W x H):250 mm x 108 mm x 42 mm
Measuring probe (Ø x H).112 mm x 260 mm
Connecting cable:
0,7 m (up to 100 m optionally available)
Weight:Display unit with measuring probe: 1,5 kg



Photon Energy Response:





Angular Energy Response:


print page Top of the page
© Sensortechnik und Elektronik Pockau GmbH | created by HMS Multimedia